Monroe-electronics Electrostatic Voltmeter - Isoprobe® - model 279L Manual do Utilizador Página 15

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 24
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 14
Page 12
7. Check the response to negative steps in a similar manner.
8. Switch to standby.
J. Noise Check:
1. Replace the oscilloscope with an rms voltmeter. Noise is typically 0.6mV rms.
K. Output Calibration:
1. Connect the rear panel output connector to a DVM
2. Adjust ZERO 1 for zero volts on the DVM and the front panel meter.
3. Slowly apply voltage to the test plate until it is precisely +300V.
4. The DVM at the output should +3.00V
0.1%. Adjust R5 if necessary to meet this condition.
5. The front panel meter should indicate +300.0
0.1% LSD.
6. Repeat steps 3 through 5 using –300v.
7. Remove the test equipment and replace the cover. This completes the adjustment procedure.
Vista de página 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 23 24

Comentários a estes Manuais

Sem comentários