Monroe-electronics Electrostatic Voltmeter - Isoprobe® - model 244A Manual do Utilizador Página 5

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 24
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 4
5
SECTION 1
SPECIFICATIONS
A. APPLICATIONS:
The Model 244A instrument permits accurate voltage measurements of
electrostatic or other high impedance sources without physical contact. It utilizes
an electrostatic chopper for low drift, and negative feedback for accuracy and
probe-to-surface spacing insensitivity.
B. FEATURES:
Integrated Solid State Circuitry
Good Accuracy
Stability
Reliability
Easy Maintenance
±3kV Input Range
Fast Speed-of-Response
Specified AC Performance
Low Noise
Digital Liquid Crystal Display
HV/1000 Recorder Output
Connectable for unlimited multiple channel operation
C. RANGE: 0 to ±3000 volts DC.
D. STATIC ACCURACY:
The recorder output accuracy referred to the input is better than 0.10% of reading
or 0-1V whichever is greater. (Probe-to-surface spacing .020”, 0.5mm).
E. SPEED-OF-RESPONSE:
10-90% probe-to-surface spacing .040" (1.02mm), gain adjusted for critical
damping.
Speed of Response
typ. max.
1kV Step .......................................... 1.5 msec ................. 2.5 msec
2kV Step .......................................... 2.0 msec ................. 3.5 msec
3kV Step .......................................... 3.0 msec ................. 4.5 msec
F. SETTLING TIME:
typical - 3.5msec. 0-99% 1kV step, 040"(1.02mm) gain adjusted for critical
damping.
15 msec. 0-99.9% 1kV step, .040"(1.02mm)
Vista de página 4
1 2 3 4 5 6 7 8 9 10 ... 23 24

Comentários a estes Manuais

Sem comentários